System Messages
Table 12-2 Momentary Messages (Continued)
Message
12-6
Possible Cause
The incident is more than 2 hours in
duration, causing the Data Card to fill.
An empty Data Card was not inserted
for the incident, causing the Data Card
to fill sooner.
The Data Card is removed during an inci-
dent.
The Data Card is inserted while the Heart-
Start XLT is on.
A Data Card other than the M3510A is
inserted.
A Data Card is not in the HeartStart XLT.
The key pressed is currently inactive (i.e.
Sync
Pacer
and
AED Mode).
Mode
is pressed while pacing pulses
are being delivered.
None. A Data Card must be inserted prior
to turning the HeartStart XLT on for the
current patient.
Use only M3510A Data Cards.
Turn the HeartStart XLT off and insert a
Data Card prior to the first event for the
patient.
Use the appropriate mode for the key.
are inactive in
Stop pacing before changing the pacing
mode.
Corrective Action
None. A new Data Card can not be
inserted during an incident.
Use one Data Card per incident/
patient to decrease the chance of the
card filling.
None. The Data Card can not be re-
inserted during an incident.
Do not remove the Data Card during
an incident.
Troubleshooting