To replace the diffusion pump fluid

To replace the diffusion pump fluid
Materials needed:
Aluminum foil, clean
Cloths, clean, lint-free (05980-60051)
Container for old diffusion pump fluid
Diffusion pump fluid, 18.5 ml (6040-0809) – 2 required
Gloves
1 Remove the diffusion pump from the MSD (page 114).
Make sure you remove the O-ring assembly from the top of the diffusion pump.
2 Cover the top of the diffusion pump with aluminum foil.
3 Heat the diffusion pump at 60 °C for 15 minutes in your GC oven.
: $ 5 1 , 1 *
7KH SXPS DQG SXPS IOXLG ZLOO EH KRW :HDU SURWHFWLYH JORYHV ZKHQ \RX UHPRYH WKH SXPS
IURP WKH RYHQ
4 Pour the old diffusion pump fluid out the top of the pump.
Even after heating, the pump fluid pours very slowly.
: $ 5 1 , 1 *
7UHDW WKH ROG SXPS IOXLG DV KD]DUGRXV ,W PD\ FRQWDLQ WUDFHV RI WR[LF FKHPLFDOV
Charred or blackened
If the diffusion pump has been heated with insufficient pump fluid (or with a large
pump fluid
air leak in the MSD), the remaining pump fluid may be severely blackened. Black-
ened pump fluid may also be baked onto the internal parts (stack) of the pump. If
so, you may have to remove the diffusion pump stack and clean its parts, and the
interior of the pump, with methylene chloride. Be very careful when reinstalling
the stack. Misalignment of stack components can seriously reduce diffusion pump
performance.
: $ 5 1 , 1 *
0HWK\OHQH FKORULGH LV D KD]DUGRXV VROYHQW :RUN LQ D IXPH KRRG DQG WDNH DOO DSSURSULDWH
SUHFDXWLRQV
116