Advertisement

Maintaining the MSD
4
To clean the ion source
8 Dry the cleaned parts in an oven at 100 ° – 150 °C for 30 minutes.
: $ 5 1 , 1 *
/HW WKHVH SDUWV FRRO EHIRUH \RX KDQGOH WKHP
Take care to avoid recontaminating cleaned and dried parts. Put on new, clean
gloves before handling the parts. Do not set the cleaned parts on a dirty surface.
Set them only on clean, lint-free cloths.