Before starting

You can perform much of the maintenance required by your MSD. For your
safety, read all of the information in this introduction before performing any
maintenance tasks.
Some parts of the MSD require regularly scheduled maintenance
Common maintenance tasks are listed in Table 4. Performing these tasks
when scheduled can reduce operating problems, prolong system life, and
reduce overall operating costs.
Keep a record of system performance (tune reports) and maintenance
operations performed. This makes it easier to identify variations from nor-
mal operation and to take corrective action.
Table 4
Maintenance schedule
Task
1
2
94
Tune the MSD
Check the foreline pump oil level
Check the calibration vial
Replace the foreline pump oil
Check the diffusion pump fluid
Clean the ion source
Check the carrier gas trap(s) on the GC
Replace the worn out parts
Lubricate sideplate or vent valve O-rings
Every 3 months for CI MSDs using ammonia reagent gas.
Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be
lubricated. Lubricating other seals can interfere with their correct function.
Every
week
1
2
Every
Every
6 months
year
As
needed