To remove the foreline gauge

To remove the foreline gauge
Materials needed:
Screwdriver, flat-blade, large (8730-0002)
1 Vent the MSD (page 52).
2 Separate the MSD from the GC (page 112).
: $ 5 1 , 1 *
7KH IRUHOLQH SXPS DQG GLIIXVLRQ SXPS PD\ VWLOO EH KRW
3 Unplug the foreline gauge cable from the foreline gauge.
4 Disconnect the foreline gauge assembly from the diffusion pump outlet.
5 Loosen the hose clamp.
6 Pull the foreline gauge assembly out of the foreline hose.
& $ 8 7 , 2 1
Be sure the MSD is vented to atmosphere before breaking the seal at the foreline gauge.
Never
126
Maintaining the MSD
vent the MSD at the pump end; use the vent valve.