To Measure Column Flow Linear Velocity - HP 5973 Hardware Manual

Mass selective detector
Hide thumbs Also See for 5973:
Table of Contents

Advertisement

2

To measure column flow linear velocity

To measure column flow linear velocity

Materials needed:
Syringe
1 Set Data Acquisition for manual injection and selected ion monitoring
(SIM) of m/z 28.
2 Click
3 Inject about 1 µl of air into the injection port and click
4 Wait until a peak elutes at m/z 28.
Note the retention time.
5 Calculate the average linear velocity.
Average linear velocity (cm/sec)
where:
L = length of the column in meters
t = retention time in seconds
Be sure to account for any pieces of column broken off. A 1-meter section missing
from a 25-meter column can yield a 4% error.
6 Use this value to verify the MSD ChemStation flow calculations (page 47).
If the numbers disagree, click the
sions.
7 To calculate the volumetric flow rate.
Volumetric flow rate (ml/min) =
where:
D = internal column diameter in millimeters
L = the column length in meters
t = the retention time in minutes
46
Operating the MSD
.
Prep Run
100 L
=
-------------- -
t
button to calibrate the column dimen-
Change
2
0.785
D
L
------------------------------
t
.
Start Run