Temperature Symptoms; Ion Source Will Not Heat Up - HP 5973 Hardware Manual

Mass selective detector
Hide thumbs Also See for 5973:
Table of Contents

Advertisement

Troubleshooting the MSD
3

Temperature symptoms

Temperature symptoms

The MSD has three heated zones:
• Ion source (
Source
in the MSD ChemStation software)
• Mass filter (
in the MSD ChemStation software)
Quad
• GC/MSD interface (
Thermal Aux #2
in the MSD ChemStation software)
Each of these heated zones has a heater cartridge and temperature sensor. The ion
source and mass filter are powered and controlled by the MSD. The GC/MSD
interface is powered and controlled by the GC.

Ion source will not heat up

• High vacuum pump is off or has not reached normal operating conditions*
• Incorrect temperature setpoint
• Ion source has not had enough time to reach temperature setpoint
• Ion source heater cartridge is not connected*
• Ion source temperature sensor is not connected*
• Ion source heater failed (burned out or shorted to ground)*
• Ion source temperature sensor failed*
• Source power cable is not connected to the side board*
• MSD electronics are not working correctly
* These will cause an error message
82