Column
Interface column nut
Column installation tool
1 to 2 mm
Interface ferrule
Septum
8 Clean the outside of the end of the column with a lint-free cloth moistened
with methanol.
9 Insert the column into the GC/MSD interface.
10 Tighten the nut 1/4 to 1/2 turn.
Check tightness after one or two heat cycles.
& $ 8 7 , 2 1
The column installation tool must be kept
the ion source. Keep it in its storage tube, and clean it by flushing with methanol after each
use.
To install a capillary column using the installation tool
to prevent contaminating the column and
clean
Installing GC Columns