10.Press the
key (below the Next prompt).
The Sample Flow Recovery status message is displayed:
Flow Rate Test
Sample Flow Recovery
Next
Wait until the progress bar indicates completion.
Flow rate test >5 % O
2
IMPORTANT (NOTE)
Refer to Operating instruction OI/AZ40-EN for
troubleshooting procedures.
To perform a flow rate test (>5 % O
1. From any Operator page, press the
Autoscroll
In Conf Alarm
The Operator menu is displayed:
Data Views
Logs
Alarm Acknowledge
Autoscroll
In Conf Alarm
Enter Configuration
2. Use the
keys to scroll to the Enter Configuration
/
menu and press the
The Access level screen is displayed:
CI/AZ40-EN Rev. A
) at the transmitter:
2
key.
CAL
key.
Access Level
Logout
Read Only
Calibrate
Advanced
Service
Back
3. Use the
keys to scroll to the Advanced level and
/
press the
key to display Advanced level menu options.
IMPORTANT (NOTE)
If passwords have been set it is necessary to enter
the correct password to enable access to the
Advanced level – refer to Operating instruction
OI/AZ40-EN for password setup details.
4. Use the
keys to scroll to the Calibrate level screen:
/
Menu
Calibrate
Exit
5. Press the
key to enter Calibrate level and display menu
options, then use the
Rate Test menu:
Calibrate
Calibration Setup
Blowback Setup
Scheduled Events
Flow Rate Test
Man. Valve Control
Back
Select
Select
keys to scroll to the Flow
/
Select
23