Specification
Power supply requirements
Supply voltage
85 to 265 V AC, 50 / 60 Hz
— transmitter: <60 W
— sensor: <60 W, <730 W (during start up) and <310 W
(when operating)
Probe insertion lengths
Dimensions in mm (in.)
Standard probe
Temperature range –20 to 650 °C (0 to 1,200 °F)
No filter
Primary filter
600 (24)
950 (37)
900 (36)
1265 (50)
1200 (48)
1550 (61)
1500 (60)
1850 (73)
1800 (72)
2150 (85)
2100 (84)
2460 (97)
High temperature probe
Temperature range –20 to 1650 °C (0 to 3000 °F)
No filter
High temperature filter
600 (24)
850 (34)
900 (36)
1250 (49)
1200 (48)
1550 (61)
Maximum probe process temperature by filter type
Standard probe
Filter type
Maximum temperature
Primary
649 °C (1200 °F)
Primary + secondary 816 °C (1500 °F)
High temperature probe
Probe length
Maximum temperature
600 (24)
1650 °C (3000 °F)
900 (36)
1370 °C (2500 °F
1200 (48)
1232 °C (2250 °F)
Process pressure range
±5 kPA (±20 in. WG)
Process connections
Standard / high temperature probes
ANSI 2 / 3 / 4 in.
DIN 80 / 100
Digital outputs
6, normally-closed 2 A at 230 V AC (30 V DC non-inductive)
Digital inputs
4, volt-free contact
EMC
Emissions and immunity
EN61326 Industrial specification
General safety
CE (EN61010)
CI/AZ40-EN Rev. A
Primary + secondary filter
1150 (45)
1465 (57)
1750 (69)
2050 (81)
2350 (93)
2660 (105)
Sales
Service
35