ABB Endura AZ40 Manual page 25

Oxygen and carbon monoxide equivalent coe analyzer
Hide thumbs Also See for Endura AZ40:
Table of Contents

Advertisement

Setting up test gases
IMPORTANT (NOTE)
Refer to Operating instruction OI/AZ40-EN for
troubleshooting procedures.
To set up test gases at the transmitter:
1. From any Operator page, press the
Autoscroll
In Conf Alarm
The Operator menu is displayed:
Data Views
Logs
Alarm Acknowledge
Autoscroll
Enter Configuration
In Conf Alarm
2. Use the
keys to scroll to the Enter Configuration
/
menu and press the
The Access level screen is displayed:
Access Level
Logout
Read Only
Calibrate
Advanced
Service
Back
CI/AZ40-EN Rev. A
key.
CAL
key.
Select
3. Use the
keys to scroll to the Advanced level and
/
press the
key to display Advanced level menu options.
IMPORTANT (NOTE)
If passwords have been set it is necessary to enter the
correct password to enable access to the Advanced
level – refer to Operating instruction OI/AZ40-EN for
password setup details.
4. Use the
keys to scroll to the Calibrate level
/
screen:
Menu
Calibrate
Exit
5. Press the
key to enter Calibrate level and display menu
options, then use the
Calibration Setup menu:
Calibrate
Manual Calibration
Manual Blowback
Calibration Setup
Blowback Setup
Scheduled Events
Flow Rate Test
Back
6. Press the
key (below the Select prompt).
The Calibration Setup screen is displayed:
Calibration Setup
CO Zero Test Gas
CO Span Test Gas
O2 Zero Test Gas
O2 Span Test Gas
Zero Gas Cal. Time
Span Gas Cal. Time
Back
Select
keys to scroll to the
/
Select
Select
25