Mechanical installation

Standard temperature probe assembly (primary and optional
filter arrangements) – all flange options
Referring to Fig. 5:
1. Apply a light coating of an anti-seize compound (suitable for
temperatures up to 200 °C [392 °F]) to both threaded ends
A
B
of probe
and to threaded end
secondary filter.
2. Thread primary filter assembly
D
filter assembly
onto probe shaft
3. Thread the probe assembly with attached filter(s) onto the
F
inch NPT port
on sensor assembly
4. Apply an anti-seize compound (suitable for temperatures up
to 200 °C [392 °F]) to the threaded end of sensor aspirator
H
.
5. Hand-tighten exhaust filter assembly
H
aspirator thread
.
Fig. 5 Standard temperature probe assembly – all flange options
CI/AZ40-EN Rev. A
C
of (optional)
E
and (optional) secondary
B
.
1
G
and tighten.