Figure 3-2 Operational Block Diagram (Maxum Ii Airless Airbath Model Shown) - Siemens Maxum II Manual

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

System Functions
3.3 Analyzer Operation
GCP
Figure 3-2
Power On
The Power Entry Control Module (PECM), in response to commands on internal bus, accepts
system primary power and provides switching and control of AC power for oven heaters and
other AC powered devices.
Sample Conditioning
Before being piped to the analyzer, the sample from the process is sent to a sample conditioner
system. The sample conditioner ensures that the process sample is compatible with the
requirements of the analyzer. That is, it assures that the phase, pressure, temperature and
flow rate to the analyzer are suitable, that the sample is filtered, that condensates are removed
and other treatments are carried out. The resultant conditioned sample is typically piped via
1/8-inch stainless steel tubing to the sample valve(s) located in the oven of the Maxum II.
14
Network
SYSCON
SNE Software Moule
DPM
DPM
(A/D)
(A/D)
FID
Feedthroughs
TCD
Oven
Operational Block Diagram (Maxum II Airless Airbath model shown)
Electronics Enclosure
2
I
C
Electronic
Pressure
Control
Module
Temperature
Control
DPM
(A/D)
Detector Compartment
FPD
Atmosphere Vents
I
2
C from SYSCON
Sample
Conditioning
Sample
Valve
Regulated Carrier Gas
Maxum edition II Analyzer General Maintenance
Solenoid
Valve
Control
Module
Oven
Heater
Control
(Mezannine)
Sample In
Manual, August 2018, A5E42019842001