General Maintenance and Troubleshooting
4.6 Alarm Codes, Descriptions, and Suggested Actions
#
Text
403
!
Configured Detector %3
Balance Failure
404
? All Methods must be in
Hold before Configuring
Detectors
405
!
Method not on tracking list:
Count = %3 Track ID = %4
406
!
Error %3 Monitoring Purge
Signal
407
? SYSCON-SNE Communi‐
cations Overload Detected
408
? Spurious detector acquisi‐
tion
409
? SNE low on Memory
420
!
Heartbeat timeout
421
!
Heartbeat lost
422
!
Cannot connect to RSP
423
!
Method %3 has more than
8 simultaneous events
424
!
Method %3 has more than
255 events
38
Description
Balance Failure from an Advance+ de‐
tector.
All methods must be in hold while chang‐
ing the configuration of any configured
detectors.
SNE processing error.
SNE or I2C error.
SNE or I2C messaging error.
SNE error.
SNE error.
MicroSAM: error
MicroSAM: error
MicroSAM: error
MicroSAM: error
MicroSAM: error
Action
Check sys_detector_cfg configuration.
Set all application in hold and wait for
cycles to complete. Then reconfigure
Advance Plus configured detector.
Contact Customer Support.
Reset unit.
Contact Customer Support.
Contact Customer Support.
Reduce the memory consumption.
Some examples of how to do this in‐
clude:
Reduce the number of peaks detected
by increasing the threshold.
Reduce the length of cycles.
Reduce the detector sampling rate.
Contact MicroSAM support for assis‐
tance.
Contact MicroSAM support for assis‐
tance.
Contact MicroSAM support for assis‐
tance.
Contact MicroSAM support for assis‐
tance.
Contact MicroSAM support for assis‐
tance.
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001