Alarm Codes, Descriptions, And Suggested Actions; Alarms 301 - 324 - Siemens Maxum II Manual

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

4.6

Alarm Codes, Descriptions, and Suggested Actions

4.6.1

Alarms 301 - 324

GCP Alarm Descriptions 301 - 324
The following tables list the alarm number (#), type (+ information, ? warning, ! error) alarm
text, description, and actions.
#
Text
301
? External Message: send
failure %3
302
?
External Message: server
lost
303
+
External Message: Rec'd
invalid communication
from unit %3
304
+
External Message: Orphan
message received from %3
305
?
External Message: Invalid
Message Length for %3
306
+
External Message: Send in‐
valid communication to unit
%3
307
?
External Message: Dupli‐
cate anlz_id %3 detected ;
setting to zero
308
?
External Message: Dupli‐
cate UNIT %3 detected;
setting loop/unit to zero
309
?
External Message:
RUD:Unit does not re‐
spond; loop/unit %3
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001
4.6 Alarm Codes, Descriptions, and Suggested Actions
Description
A message was received by the device
from itself; or source of message can't be
identified; or Gateway is too busy or com‐
munication was disrupted between the
GC and the message handler.
System Error
Message was received from Advance
DataHiway unit that had previously
broadcast with no slots or slot is out of
range.
An Advance DataHiway external
PANDSP message was received with no
matching PANKEY; or an ATTACH was
received with no matching RATCH; or an
internal timeout was generated for non-
existent message.
Advance DataHiway Results, print, HAE,
or Service Panel messages received
from GC that have no length.
Message from GC is directed to an Ad‐
vance DataHiway unit that has no slots,
an invalid range of slots, or no UID has
ever been received.
An Advance DataHiway ZIP message
was received from another Advance Da‐
taHiway unit, or another GC has broad‐
cast with the same analyzer num‐
ber.Check other units on network.
An Advance DataHiway SLEEP mes‐
sage has been received. Another Ad‐
vance DataHiway unit has broadcast
with the same loop/unit.
Occurs when Advance DataHiway loop/
unit does not respond to a RUD message
- originates from I/O, Host, or Printer ta‐
ble in GC.
General Maintenance and Troubleshooting
Action
Reset Gateway or SYSCON.
Contact Customer Support.
Reset Advance DataHiway unit.
Ignore or reset SYSCON.
Check database set up for these mes‐
sages.
Check Advance DataHiway unit.
Check other units on network.
Check other units on Advance DataHi‐
way and correct loop/unit of GC.
Remove extraneous references to non-
existent units. Check Advance Data‐
Hiway connection.