Siemens Maxum II Manual page 76

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

General Maintenance and Troubleshooting
4.6 Alarm Codes, Descriptions, and Suggested Actions
#
Text
4028
? Detector channel unknown
error occurred on %4
4029,
? Zero Correction failure on
4030
[first/lower/right; second/
lower/left] channel of %4
4031
!
Flame out on %4
4032
? Purge Loss on %4
4033
!
Zero Correction Out of Lim‐
its on %4
4120
? Pulse DO Within a Pulse
DO is Not Allowed on %4
4121
+ Pulse DO Aborted on %4
4122
!
Method Conflict: Concur‐
rent Chromatogram Chan‐
nel Acquisition on %4
4123
!
Unsupported detector sam‐
pling rate on %4
4124
!
Run-time Operational Con‐
flict on %4
76
Description
Undeterminable error occurred on detec‐
tor channel circuit. This error may also be
a secondary error caused by an overflow
on an FID DPM.
Should occur only with a revision 2 TCD
DPM or revision 3 baseboard (PIC firm‐
ware revision 2.006 and up ). Or, the zero
correction value exceeded the maximum
allowed value for that particular board.  
This is a balance failure situation.
Flame out detected on an Advance+ FID
or FPD Board.
Loss of purge pressure has been detec‐
ted. Possible causes include low pres‐
sure for supply air, open door, and failed
gaskets. Depending on the environment
classification where the analyzer is instal‐
led, this alarm may indicate an unsafe
condition that requires immediate action
to correct.
The value of the zero correction (soft‐
ware balance of the detector) has excee‐
ded the balance limits.
A pulse DO was requested by the SY‐
SCON while a pulse DO was already in
progress.  The latest requested pulse
DO was ignored.
A pulse DO was aborted as a result of a
standard (non-pulsed) DO command.  
 The non-pulsed DO command has
precedence.
It is not allowed to use the same detector
channel for more than one chromato‐
gram acquisition at a time, whether within
a method or across methods.  The new‐
est request was cancelled.
A request was made to run the specified
detector channel at a standard rate but
the DPM hardware does not support that
rate. The method file is likely corrupt. It
may be necessary to rebuild the method.
An operation was requested at a time
when it is not allowed. e.g.: GainSelect
VDO: A gain select was initiated before
the previous GainSelect sequence was
completed.
Action
Reset. If error repeats, replace DPM.
Eliminate the cause of the balance fail‐
ure:
On TCD detector, replace the detector
bead or filament.
On FID detector, clean or replace the
FID.
Ignite flame on unit. If flame is burning,
board may be defective.
Investigate and correct cause of pres‐
sure loss.
Confirm that the corresponding balance
limit values are not too small. See the
maintenance instructions for the specif‐
ic detector type.
Wait until the on-going pulse has com‐
pleted before requesting another pulse
DO.
Informational message. No action nec‐
essary.
Verify that the method does not contain
2 concurrent uses of the same detector
channel hardware. Verify that the meth‐
ods running in separate applications
are not erroneously reusing the same
detector channels.
Contact Customer Support for assis‐
tance.
Revise the method to remove the con‐
flicting events.
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001