Pecm Leds - Siemens Maxum II Manual

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

4.1.2.3

PECM LEDs

The LEDs on the PECM board can help with troubleshooting problems. There are two sets of
LEDs: one on each side of the front board as shown in the diagram to the right. The bottom
set of three LEDs is the same as used on other boards. The left set is for the PECM software
and the right set is for the temperature controller software. Each will go through a separate
display on power up and normal operation. The top two sets of four LEDs each are for the
operation of each of the two temperature controllers on the front board. Heater 1 is on the left
and heater 2 is on the right side of the board.
PECM LED Interpretation
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001
LEFT Heater Status
Heater 1 Air Pressure
Heater 1 Power Activate
Heater 1 Temp Limit
Heater 1 Overtemp
PECM Status
PECM LEDs
Normal
Fault
Warning
General Maintenance and Troubleshooting
4.1 Analyzer Status Indicators
RIGHT Heater status
Heater 2 Air Pressure
Heater 2 Power Activate
Heater 2 Temp Limit
Heater 2 Overtemp
Normal
Normal
Fault
Fault
Warning
Warning
Temperature
Controller
Status
25