Alarms 801 - 999 - Siemens Maxum II Manual

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

#
Text
732
!
IO: underflow or lower fail‐
safe condition detected for
%3
733
!
IO: overflow or upper fail‐
safe condition detected for
%3
734
!
%3 Process not communi‐
cating
735
? IO: AO value was clamped
for %3 to %4
736
? Untrusted Connection Re‐
fused from %3
737
? System Time is Incorrect
4.6.7

Alarms 801 - 999

GCP Alarm Descriptions 801 - 999
The following tables list the alarm number (#), type (+ information, ? warning, ! error) alarm
text, description, and actions.
#
Text
801
? System Error %3 in File %4
line %5
802
? Error %3 opening Flash
File %4
803
? Error %3 closing Flash File System Error
804
? Error %3 reading Flash File System Error
805
? Error %3 writing Flash File
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001
4.6 Alarm Codes, Descriptions, and Suggested Actions
Description
The firmware is reporting an under-range
error for analog NAMUR data types 6 or
8.
The firmware is reporting an under-range
error for analog NAMUR data types 7 or
8.
Occurs when ADHMaxumD or I2CDB
connection is not present.
Indicates clamping of AO value. Informa‐
tional message.
A connection was refused from an exter‐
nal client due to the trusted-connection
function.
System Time is Incorrect
Description
System Error
System Error
System Error
General Maintenance and Troubleshooting
Action
Check the wiring for the secondary de‐
vice being used. Verify proper ranges
for the I/O.
Check the wiring for the secondary de‐
vice being used. Verify proper ranges
for the I/O.
Contact Customer Support.
No action necessary.
Trusted Connections are defined on the
HMI. Please contact your site's admin‐
istrator or Siemens Customer Service.
Set Analyzer Time or set TimeServer IP
Address as described in "Regional Set‐
tings" Help in GCP.
Action
Contact Customer Support.
Reset the device. If the error persists,
replace the Flash/SRAM module.
Reset the device. If the error persists,
replace the Flash/SRAM module.
Reset the device. If the error persists,
replace the Flash/SRAM module.
Reset the device. If the error persists,
replace the Flash/SRAM module.
45