General Maintenance and Troubleshooting
4.6 Alarm Codes, Descriptions, and Suggested Actions
#
Text
711
+ Database: %3
712
!
%6 Start Ver: %3 - %4 on
%5
713
+ System backed up
714
+ All alarms cleared
715
+ Database Build
716
? DB: Invalid Sourcekey or
SourceAttribute for Stat‐
Mon table: ID %3
717
? DB: Calibration : margin
check/reports are invalid
for curve type
718
? Validation failed for %3
719
+ Database: SNE reset re‐
quested
720
? Database: Reprocess dur‐
ing Run not allowed
721
+ Application is in service
722
? Reference Component not
found; Component %3 ,
Reference %4
723
!
Database: Method ID %3
not found
724
? Calibration or Validation
failed; application in hold
725
? DB: Method %3 halted
726
!
DB: Slave application
stream is invalid %3
727
? DB: Master app autocali‐
bration completed before
Slave app %3
728
? DB: application cannot au‐
tocalibrate
729
!
DB: message processing
timeout for message %3;
attempting recovery
730
!
DB: Unnamed peak pro‐
cessing exceeded (2000)
for channel %3
731
!
CAN initialization failure for
application %3
44
Description
System error
SYSCON has been reset.
Informational message.
System error
System error
System error
Occurs when StatMon table is not prop‐
erly configured.
Occurs when a margin is set in EZChrom
for a curve type that does not support
margin checking.
Validation has failed.
SNE is indicating that it has been reset
from a database request.
Reprocess button pushed while applica‐
tion in run.
Informational message.
Quantra alarm.
Method ID is in sequence, but not in
method table.
Calibration or validation was requested
while one was already running.
Application was halted from the HMI,
CIM display, or a MaxBasic program. In‐
formational message.
A stream ID in slave does not line up with
master application stream ID (new re‐
quirement for version 4.0).
Master autocalibration sequence is fin‐
ishing before slave autocalibration.
Autocalibration was requested on appli‐
cation that is not defined for autocalibra‐
tion.
System error
Processing for unnamed peaks must not
exceed 2000 peaks for a channel. Ex‐
cess peaks were discarded.
Application does not start until CAN
cards required by the application are ini‐
tialized. The application starts regardless
of the error after 30 seconds.
Action
Contact Customer Support.
No action necessary.
Contact Customer Support.
Contact Customer Support.
Contact Customer Support.
Consult documentation.
Remove margin or change curve type.
Check report or validation results
screen on the HMI or CIM Display.
May indicate a communication over‐
load.
Only push reprocess button when ap‐
plication is in hold.
No action is necessary.
Contact Customer Support
Check method table.
Check timing of calibration or validation
events.
No action necessary.
Check master and slave sequences.
Check length of sequences. Master au‐
tocalibration sequence must finish after
slave autocalibration.
Check autocal attribute in application
table.
Contact Customer Support.
Increase the EZChrom threshold value
to reduce the number of peaks.
Verify that all application I/Os are prop‐
erly initialized. Remove any I/O that de‐
pends on a CAN card that is not present.
Maxum edition II Analyzer General Maintenance