Alarms 1002 - 1128 - Siemens Maxum II Manual

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

#
Text
997
+ %3
998
? %3
999
!
%3
4.6.8

Alarms 1002 - 1128

GCP Alarm Descriptions 1002 - 1128 SNE Common Module Errors
The following tables list the alarm number (#), type (+ information, ? warning, ! error) alarm
text, description, and actions.
#
Text
1002
!
ID Key Not Connected on
%4
1003
!
ID Key Change on %4
1004
!
EEPROM Bad Checksum
on %4
1005
? Temp Diag Error on %4
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001
4.6 Alarm Codes, Descriptions, and Suggested Actions
Description
General alarm used for information.
Used most often by MaxBasic programs.
General warning alarm used for informa‐
tion. Used most often by MaxBasic pro‐
grams.
General fault alarm used for information.
Used most often by MaxBasic programs.
Description
All PICs: The module location ID connec‐
tor is disconnected or set to 0. This is an
abnormal condition; the module may not
be operational.
All PICs: The module location ID value
was changed while the module was op‐
erating. This is a transient error that cau‐
ses the module to automatically reset.
The module can then be addressed and
operated at the new location ID.
All PICs: A checksum error was detected
in the module EEPROM. The firmware
will still use all the information that it can
read from the EEPROM. However, the
module may not operate normally.
All PICs: The on-board temperature sen‐
sor (LM-75) diagnostic failed. This alarm
indicates that the ability of the board to
detect a module overheat (alarm #1044)
may be compromised. This alarm is may
happen occasionally following a board
reset.
This alarm is in no way related to and
should not be confused with the Over‐
temp Shutdown related to heater temper‐
ature controls.
General Maintenance and Troubleshooting
Action
No action necessary.
Immediately report runtime errors to
Customer Support (please make care‐
ful note of the alarm message).
Immediately report runtime errors to
Customer Support (please make care‐
ful note of the alarm message).
Action
Verify that the location ID connector is
in good condition and connected prop‐
erly.
Verify that the location ID connector is
in good condition and connected prop‐
erly.
Check for intermittent connection.
Cycle power.
If the error repeats, replace the module.
No action is required unless the error
happens every time the board is reset.
For these repetitive errors replace the
module.
47