Disinfecting The System - Toshiba Aplio 500 TUS-A500 Operation Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

20.3.2 Disinfecting the system

This subsection describes the procedures and precautions for disinfecting the system.
For disinfection of transducers, refer to the operation manuals provided with the
transducers.
DANGER:
WARNING
CAUTION:
CAUTION: 1. When chemical disinfections are repeated, the surface of the system is
2. If any abnormality related to functions of the product is observed after
After disinfecting the system, ventilate the room fully before
turning the system ON.
If flammable and explosive gas remains near or in the system,
it may cause a fire or explosion.
: 1. Before disinfecting the system, be sure to disconnect the
system power cable plug from the outlet of the facility.
If the system is defective, an electric shock may occur.
2. Do not spill or spray liquids such as water onto the
system or peripheral units. If a liquid such as water
enters the system or peripheral units, an electric shock
may occur.
1. Observe the following precautions to prevent infection.
 Wear sterile protective gloves.
 Wear new sterile protective gloves each time disinfection
is performed.
2. Do not disinfect the system using a method or chemical
other than those specified in this manual. The system may
not be disinfected properly, or the surface of the system may
deteriorate or become discolored.
3. After chemical disinfection, dry the surface of the system
fully.
4. The efficacy of the disinfectants is not guaranteed by
TOSHIBA. Contact the manufacturer of the relevant
chemical for the efficacy of each disinfection agent.
5. Determine the effectiveness of the chemical based on the
criteria (effective period, number of times of use,
discoloration of the chemical, results of testing using
effectiveness test kits, etc.) described in the documentation
provided by the manufacturer of the chemical.
changed gradually.
disinfection, stop using the product and contact your TOSHIBA service
representative for repair.
No. 2B771-004EN*M