Starting An Examination - Toshiba Aplio 500 TUS-A500 Operation Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

11. Starting an Examination

CAUTION:
NOTE: 1. Be sure to register and save the patient ID before starting an examination
for a new patient. If measurement is performed for a new patient for whom
the patient ID has not been registered or saved, the measurement results
and other data cannot be saved in the system.
2. Periodically back up the saved data (image data and exam history) to
external storage media. Data stored in the system may be deleted as a
result of an operational error or an accident (problems that occur in
personal computers can also occur in this system).
3. Delete unnecessary data regularly. If the available storage space is low,
new data may not be saved.
4. Before deleting data, confirm the date and the patient ID, patient name,
etc. to ensure that incorrect data is not deleted.
5. If patient data or exam data is to be modified, confirm that the modifications
are correct before saving them. Once the data is saved, the previous data
cannot be restored.
6. If the rows of the History data in the patient data entry screen are switched,
turn ON the system again.
7. When the patient information is registered and
selected, the live image is normally displayed within 10 seconds.
However, if relatively high-load processing (such as saving, loading,
copying, or conversion of a large amount of data) is performed immediately
before this, it may take approximately 30 seconds to display the live image.
If the live image is not displayed after 30 seconds, an error message is
displayed. Reboot the system.
1. Before starting an examination for a new patient, confirm
that the patient ID matches the patient to be examined. If
images are recorded with an incorrect patient ID, the data
may be mixed up with that for another patient, resulting in
incorrect diagnosis.
2. Enter the height and the weight correctly. Otherwise, the
BSA may be calculated incorrectly.
3. Check the date format before entering the date of birth.
Otherwise, the age may be calculated incorrectly.
No. 2B771-004EN*M
11-1
is then

Advertisement

Table of Contents
loading

Table of Contents