Toshiba Aplio 500 TUS-A500 Operation Manual page 83

Diagnostic ultrasound system
Table of Contents

Advertisement

NOTE: Changing the lead type by changing the reference signal cable connections without
using the ECG Lead function
1. The relationships between the lead wire colors of the reference signal cable and the
electrode positions are described below.
For the USA
Electrode positions (lead I)
To use lead II or lead III, connect the lead wires of the reference signal cable to the
electrodes as shown below.
2. This system is designed to use ECG lead I. To use lead II or lead III, the lead colors
and the connection destinations differ from the standard arrangement.
To display both the ECG waveform and the respiratory waveform (using the
reference signal cable), use lead I.
Three electrode leads
Electrode positions
Lead color
Symbol
(clip color)
RA
White
LA
Black
LL
Red
No. 2B771-004EN*M
12-3
Application position
Right infraclavicular fossa
Left infraclavicular fossa
Lowest rib on the left
anterior axillary line