4. Use Conditions
4.1
Power and Environmental Requirements
Item
Power
Line voltage
Line frequency
Power consumption
Operating
Ambient temperature
environmental
Relative humidity
conditions
Atmospheric
pressure
Storage and
Ambient temperature
transportation
Relative humidity
conditions
Atmospheric
pressure
Patient environment
Specifications
120 VAC 10%
USA
220 to 240 VAC 10%
Europe
110 to 120 VAC 10%
Other 1
220 to 240 VAC 10%
Other 2
50 Hz to 60 Hz
USA
1440 VA
Europe
1500 VA
Other 1
1440 VA
Other 2
1500 VA
10°C to 35°C
35% to 80% (no condensation)
700 hPa to 1060 hPa
-10°C to 50°C
30% to 80% (no condensation)
50% or less if the ambient temperature exceeds
40°C
700 hPa to 1060 hPa
This system is designed to be used in the
environment specified in the figure below.
No. 2B771-004EN*M
4-1