Turning On The Power - Toshiba Aplio 500 TUS-A500 Operation Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

9.2

Turning ON the Power

CAUTION: 1. Do not disconnect the power plug while the system is starting up.
2. If either of the following phenomena occurs, press and hold down
NOTE: 1. If the power is turned ON while a USB device, such as a flash drive, is
connected to one of the five external USB connectors, the system may not
start normally. Disconnect any USB device(s) before turning the power
ON.
2. When turning ON the power, confirm that the [STANDBY] LED is lit and
then press
3. Do not operate the operating panel switches or keyboard while the system
is starting up. The system may not start up normally.
4. After the main power has been turned OFF on the power panel, wait for at
least for 15 seconds before turning the power ON again.
If the main power switch is turned ON immediately after the power has
been turned OFF, the system may malfunction.
Doing so may cause the system to malfunction.
for 5 seconds or more to turn OFF the power of the system.
 The startup screen is not displayed after waiting for 30 seconds.
 The patient registration screen is not displayed after waiting for
10 minutes.
If the power is not turned ON after holding down
more, turn OFF the main power on the power panel.
Do not turn OFF the power in this manner during normal operation.
Doing so may cause the system to malfunction.
.