Performing Synchronized Cardioversion
Performing Cardioversion
If the defibrillator does not charge, refer to Chapter 7, Troubleshooting.
Applying the Paddles
1
Do not allow paste to accumulate on the hands, the paddle
WARNING
handles, or the paddle electrodes on the chest to avoid the risk of
electrical shock or burns.
2
3
4
5
6
4-4
After the shock, the ECG strip will indicate whether the 100 joule test shock
passed or failed.
Prepare the paddles by performing the following steps.
a.
Remove paddles from their holders by grasping the handles and lifting
straight up.
b.
Apply a liberal amount of Redux® paste to the electrode surface on each
paddle or use defibrillator pads.
c.
Gently rub the electrode surfaces together to evenly distribute the applied
paste.
Apply the paddles to the chest as follows.
a.
Place the left (Sternum) paddle to the right of the sternum just below the
clavicle.
b.
Place the right (Apex) paddle on the chest just below and to the left of the
left nipple, in the anterior-axillary line.
Rub the paddles slightly against the skin to increase the paddle-to-patient con-
tact.
Apply 10 to 12 kg (22 - 25 lb) of pressure per paddle.
Press the Charge button on either the right (Apex) paddle or on the instrument
front panel.
Wait for the Charge Done indicators.