Troubleshooting

Operational Checks

Keep hands clear of the paddle electrode edges. Use your thumbs
WARNING
to depress the Shock buttons on the paddle handles.
11 Grasp the paddle handles, and without removing the paddles from their holders,
12 Press
13 Verify that the messages SYNC and USE LEADS appear on the display.
14 Press
15 Turn the Energy Select control to Off (Standby).
The defibrillator is ready for use if it passes the above checklist.
A more extensive test of defibrillator/monitor functionality can be performed using
the diagnostic service mode described in the M1722/M1723 CodeMaster XL Series
Defibrillator/Monitor Service Manual.
Operational Checks
7
These checks are intended to briefly verify the proper operation of the CodeMaster
XL+ defibrillator/monitor. Regularly perform a test routine incorporating the
following checks along with visual inspection of all cables, paddles, and controls.

Every Shift

Perform the following checks every shift.
7-8
press both Shock buttons simultaneously. A brief automatic recorder run prints
the delivered energy test report.
Sync
to place the instrument in Sync mode.
Lead
once to select Lead I, and verify that the message USE
LEADS no longer appears. The message LEADS OFF should now appear, indi-
cating that one or more leads are not connected.