Troubleshooting
Operational Checks
o
Verify that the instrument is connected to AC power, and that the
BATT
and
o
Check for adequate thermal paper in the recorder.
o
Check for ECG leads, electrodes, and adequate REDUX® electrolyte paste or
defibrillator pads.

Every Day

o
Visually check AC power cord for wear.
o
Visually check the patient cables, paddles, cables, and pads adaptor cables for
wear, insulation nicks, and other damage.
o
Ensure that the
BATT
plugged in and the
broken wire.
o
Check that the
BATT
unplugged.
o
Perform the Delivered Energy and Shock Button Functional Test, which
follows.
o
If your instrument has the external pacing option, perform the Quick Pacer
Functionality Test as described in Quick Pacer Functionality Test on page 7-
11.
Delivered Energy and Shock Button Functional Test
To check the instrument with the paddles, perform the following steps.
1
Turn the Energy Select control to the 100 joules position.
1
Verify that the adult paddle electrodes are installed.
2
Push the paddles completely into their holders (Apex paddle in right pocket,
Sternum in left) and press either Charge button. Wait for the Charge Done indi-
cators.
AC
lights are on.
and
AC
AC
light is not on, the power cord may have a
and
AC
lights are on. If the unit is
lights go off when the unit is
7-9