HP CodeMaster XL+ M1722A User Manual page 88

Defibrillator/monitor
Table of Contents

Advertisement

Installation and Setup
Setup
1
Depress
Sync
HR
(Standby) to Monitor On. A menu will appear with the following choices.
CALIBRATE DEFIB
SETUP MENU 1
SETUP MENU 2
RESTORE FACTORY SETTINGS
PRINT LOG
TEST DEFIB
TEST ECG
TEST CRT
TEST RECORDER
TEST CONTROLS
TEST INDICATORS
TEST BATTERY
TEST PACER
2
Select SETUP MENU 1 by pressing
SETUP MENU 1
on
3
Press
Lead
. SETUP MENU 1 will appear with the current setup values
displayed.
ECG Size
4
Press
you wish to change.
5
Press
Lead
.
ECG Size
6
Press
this parameter.
7
When the choice you want is displayed, press
ECG Size
8
Press
value you wish to change.
9
Repeat steps 5 through 8 until you are finished configuring the settings in setup
menu 1.
10 Depress both sides of the
nostic menu.
A-8
while turning the Energy Select control from Off
ECG Size
.
or
until the highlight appears on the value
ECG Size
or
to scroll through the available choices for
ECG Size
or
until the highlight appears on the next
ECG Size
key at once to return to the setup/diag-
ECG Size
until the highlight appears
Lead
to set your choice.