Pacing (Optional)

Using the Pacer

The CodeMaster XL+ will pace on battery power alone. However, whenever
possible, plug the CodeMaster XL+ into AC power while pacing.
Using the Pacer
5
To use the pacer, perform the following steps.
1
Apply the pads (M1749A) as instructed on the package.
2
Attach monitoring electrodes as instructed in Using Leads to Monitor on page
3-1.
3
Attach the patient cable to the CodeMaster XL+'s ECG Input connector.
4
Attach the patient cable leads to the monitoring electrodes.
5
Attach the pads adaptor cable (M1750A/B) to the defibrillator output connector.
Pull the latch connector toward the front of the defibrillator to lock the connec-
tor in place.
6
Attach the pads to the pads adaptor cable and turn the twist lock.
7
Turn the Energy Select control to the Monitor On position.
If the message NO PADDLES is displayed, check that the pads adaptor cable
m
connector is properly seated and latched.
If the message PADS OFF is displayed, check the pads connection to the patient
m
and to the pads adaptor cable.
8
Press
Pacer
to turn the pacer on. Pacer parameters will now be displayed at
the bottom of the display (PACER STOP, DEMAND MODE, 70 PPM 30
MA). The rate (ppm) and output (mA) settings for when the pacer is turned on
can be selected in setup menu 1. The original rate and output settings from the
factory are 70 ppm and 30 mA. The pacer is always in Demand mode when it is
turned on.