Preparing The Leads For Monitoring - HP CodeMaster XL+ M1722A User Manual

Defibrillator/monitor
Table of Contents

Advertisement

Monitoring
Using Leads to Monitor
Table 3-1
Cardiac Monitoring Configurations
For this ECG
source...
PADS
PADDLES

Preparing the Leads for Monitoring

The CodeMaster XL+ can be configured to use either a 3-wire or a 5-wire patient
cable. Use setup menu 2 as described in Appendix A to select the patient cable type
(3-wire, 5-wire).
3-Wire Patient Cable
Table 3-2, "3-Wire Electrode Placement," on page 3-2 describes typical lead-wire
placement using the 3-wire patient cable. Table 3-3, "Lead Formation," on page 3-3
shows how the individual leads are formed using the individual leadwires.
Table 3-2
3-Wire Electrode Placement
Electrode
RA/White
LA/Black
LL/Red
3-2
Use this cable type...
Pads adaptor cable:
M1750A/B
Standard with instrument
M1746A
Placement
Near right midclavicular line, directly below the clavicle.
Near the left midclavicular line, directly below the clavicle.
Below the left pectoral muscle on the left midclavicular line.
In this monitoring
Electrode P/N
application...
ECG Normal
M1749A
Synchronized
Cardioversion
You can use these
pads for pacing, how-
ever you must select
LEADS as the moni-
tor source during pac-
ing.
Emergency ECG
N/A
Monitoring

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Codemaster xl+ m1722b

Table of Contents