Philips Respironics E30 User Manual
1.3.2
Cautions
A caution indicates the possibility of damage to the device.
Electrostatic
Do not use antistatic or conductive hoses or conductive patient tubing with the
Discharge (ESD)
device.
Pins of connectors marked with the ESD warning symbol shall not be touched
and connections shall not be made without special precautions. Precautionary
procedures include methods to prevent build-up of electrostatic charge (e.g., air
conditioning, humidification, conductive floor coverings, non-synthetic clothing),
discharging one's body to the frame of the equipment or system or to earth. It is
recommended that all individuals that will handle this device understand these
precautionary procedures at a minimum as part of their training.
Before operating the device, ensure that the SD card/filter access door and the
modem access door are both closed whenever any of the accessories such as
the Link Module are not installed. Refer to the instructions that came with your
accessory.
Condensation
Condensation may damage the device. If the device has been exposed to either
very hot or very cold temperatures, allow it to adjust to room temperature (operating
temperature) before starting therapy. Do not operate the device outside of the
operating temperature range shown in the Specifications chapter later in this manual.
Device Filters
If not using the O
blue pollen filter is required for proper operation.
Dirty inlet filters may cause high operating temperatures that may affect device
performance. Regularly examine the inlet filters as needed for integrity and
cleanliness.
Never install a wet filter into the device. You must ensure sufficient drying time for the
cleaned filter.
Extension cords
Do not use extension cords with this device.
Device placement
Do not place the device in or on any container that can collect or hold water.
Do not place the device directly onto carpet, fabric, or other flammable materials.
Do not plug the device into an outlet controlled by a wall switch.
Cleaning
Do not immerse the device or allow any liquid to enter the enclosure or the inlet filter.
Do not steam autoclave the ventilator. Doing so will destroy the ventilator.
Do not use harsh detergents, abrasive cleaners, or brushes to clean the ventilator
system.