Viral bacteria filter
A viral bacteria filter must be installed in the patient circuit (between outlet of device and patient) to prevent
contamination.
Loss of Power
The external alarm module or Uninterruptible Power Supply (UPS) must be used. If your device comes with
an external alarm module, the device should be plugged into the facility's emergency backup power outlet
(orange outlet), or a UPS should be used to provide battery back-up in the event of loss of AC mains power.
Alarm indicators
The high priority and medium priority alarms have the same auditory indication. These two alarm priorities
are differentiated by visual indications. Red signals a high priority alarm and yellow signals a medium priority
alarm. Please refer to Chapter 4 for additional alarm information.
Humidification
If the device is used invasively, a separate external humidifer is required.