Philips Respironics E30 User Manual
7.1.6 Electrical
AC Power Consumption (with 80W power supply)
Note: Power supply is part of the medical electrical
equipment.
DC Power Consumption
Fuses
7.1.7
Intake port filters
Pollen Filter
Ultra-fine Filter
7.1.8 Control accuracy
Parameter
Range
IPAP
4 - 30 cmH
EPAP
4 - 25 cmH
CPAP
4 - 20 cmH
Breath rate
0 to 30 BPM
Inspiration time
0.5 to 3 seconds
*Pressure measured at the patient connection port with or without the integrated heated humidifer (no patient flow).
7.1.9 Pressure accuracy
Pressure increments: 4.0 to 30.0 cmH
Maximum static pressure accuracy, according to ISO 80601-2-70:2015:
Parameter
10 cmH
O*
2
Static pressure accuracy has a measurement uncertainty of 3.7%
Maximum dynamic pressure variation, according to ISO 80601-2-70:2015:
Pressure
< 10 cmH
O
2
≥ 10.0 to 20 cmH
O
2
Dynamic pressure accuracy has a measurement uncertainty of 4.3%
50
Specifications
100% Polyester
88% Efficient @ 7-10 micron size
Blended Synthetic Fiber
95% Efficient @ 0.5-0.7 micron size
Accuracy
O
± 2.5 cmH
2
O
± 2.5 cmH
2
O
± 2.5 cmH
2
greater of ± 1 BPM or ±10% of setting
± (10% of setting + 0.1 second)
O (in 0.5 cmH
2
Static Accuracy
± 3.0 cmH
O
2
10 BPM
± 0.4 cmH
O
2
± 0.5 cmH
O
2
100 – 240 VAC, 50/60 Hz, 2.0-1.0 A
12 VDC, 6.67 A
There are no user-replaceable fuses.
O*
2
O*
2
O*
2
O increments)
2
15 BPM
± 0.5 cmH
O
2
± 0.8 cmH
O
2
20 BPM
± 0.8 cmH
O
2
± 1.0 cmH
O
2