Error Handling for Average Measurements - AVE
2PATH ERROR HANDLING
The
location in the same pipe to improve accuracy and the meter is operated in
Model XGS868i performs error handling only if both channels are in error. If this function is disabled, error handling
occurs when either channel goes into error.
3. Double-click Yes to enable Two-path Error Handling, or on No to disable this function.
Specific responses of the display and the totalizer to the two-path error handling option available at the above prompt
are listed in Table 25.
Option
Displays the average of CH1 and CH2, regardless
NO
of the error state of either channel.
1. If one channel is in error, the other channel's
value is displayed as the average.
YES
2. If both channels are in error, the last average
reading is held.
After responding to the above prompt, the meter returns to the
DigitalFlow™ XGS868i Programming Manual
option is intended for applications where two sets of transducers are installed in the same
Table 25: 2-Path Error Response Options
Display Response
Appendix C. Programming the XGS868i Using PanaView™
AVE
mode. With this function enabled, the
Totalizer Response
Outputs the average of CH1 and CH2 totals,
regardless of the error state of either channel.
1. If one channel is in error, totalizing continues.
2. If both channels are in error, totalizing stops.
Input/Output
option menu.
115