1.9.2h Frequency Outputs (cont.)
Setting Up the Frequency
1. Scroll to the desired Channel option and press [Enter].
2. Scroll to the desired Measurement parameter and press [Enter]. See Table 9 on page 26 for a description of the
available options. Refer to page 3-2 in the Service Manual for a description of diagnostic parameters.
Note: The measurement units that appear in these prompts are those selected in the
this section.
3. Use the arrow keys to enter a value for the Low end of the frequency output range and press [Enter].
Note: A complete pulse consists of equal amounts of ON and OFF times. Choose a value that is compatible with the
frequency counter to be used.
4. Use the arrow keys to enter a value for the High end of the frequency output range and press [Enter].
5. Use the arrow keys to enter a value between 1 and 10,000 for the Frequency at full scale and press [Enter].
Procedure Options
After completing the above steps, the XGS868i returns to the I/O window. Do one of the following:
•
To continue programming, refer to Appendix A, Menu Maps, to navigate to the desired menu.
To leave the Keypad Program, press [Escape] four times.
•
DigitalFlow™ XGS868i Programming Manual
Chapter 1. Programming Site Data
Global-System
window earlier in