Chapter 1. Programming Site Data

1.8.4 Using Advanced Features

This option enables you to access the more advanced features of the meter. In this option you can do the following:
enter a table of K factors - to compensate for non-linear flow rates (see below).
enable mass flow - calculated for static fluid density (see the next page).
1.8.4a
Entering K Factors
Use this option to enter a table of K factors. K factors are used to create a curve for the flow range (based on velocity)
that compensates for non-linear flow rates. The meter accepts from 2 to 20 data pairs. Complete the following steps to
enter multiple K factors for velocity values:
Note: The K factors are supplied by the factory; without them the K factor table cannot be edited.
1. In the Keypad Program, scroll to PROG and press [Enter].
2. In the PROG menu, scroll to CH1 or CH2 and press [Enter].
3. In the Channel PROGRAM menu, scroll to Setup and press [Enter].
4. Scroll to ADVAN. Press [Enter].
5. Scroll to MULTK, and press [Enter].
6. Scroll to ON to activate, or to OFF to deactivate multiple K factors, and press [Enter].
Note: If you selected OFF, go to Procedure Options on the next page.
7. Scroll to the Custom Type of K factor desired: CstV (velocity) or CstR (Reynolds). Press [Enter].
If you selected CstR, the meter asks for the Kinematic Viscosity. Use the arrow keys to enter the desired value,
and press [Enter].
8. Scroll to Yes to Edit Table, or to No to retain the current values, and press [Enter].
Note: If you scroll to No, go to Procedure Options on the next page.
1.8.4b
Editing K Factors
1. Use the arrow keys to enter the number of K factors (2 to 20) in the table and press [Enter].
2. Enter the velocity (or Reynolds) value for K factor number "X" and press [Enter].
Note: When editing the K factor table, the velocities must be entered in increasing order.
3. Enter the K factor corresponding to velocity number "X" (0.333 to 3.0) and press [Enter].
4. Repeat Steps 2 and 3 for each pair of values.
18