Option Card Analog Inputs
Complete the following steps to set up the analog inputs of an option card installed in
page 142):
1. Double-click on the desired Input (A, B, C or D).
Note: The setup of input A is used as an example. Identical procedures would be used to set up the remaining inputs.
2. Enter a Label of up to eight characters for input A and click on [Next Item].
3. Double-click on Off to disable Input A and return to step 1, or on Temp or Spec to designate it as a temperature or
special input.
OFF
OFF
If
was selected to disable input A, the meter returns to the Input prompt in step 1. Double-click on another input, or
click [Exit Page] to move to another menu.
TEMP or PRESR
TEMP
PRESR
If
or
was selected to set up input A as a live temperature or pressure input, complete steps 4 and 5 below:
4. Enter a value for the Base (low) end of the analog input range and click [Next Item].
5. Enter a value for the Full Scale (high) end of the analog input range and click [Next Item].
After responding to the above prompt, the meter returns to step 1. Click [Exit Page] to proceed to another menu or
option.
SPEC
SPEC
If
was selected to set up input A as a live special input, complete steps 6 through 10 below:
6. Enter a Name for input A and click [Next Item].
7. Enter the Units of measurement for input A and click [Next Item].
8. Enter a temperature value for the Base (low) end of the analog input range and click [Next Item].
9. Enter a temperature value for the Full Scale (high) end of the analog input range and click [Next Item].
10. After responding to the above prompt, the meter returns to step 1. Click [Exit Page] to proceed to another menu or
option.
DigitalFlow™ XGS868i Programming Manual
Appendix C. Programming the XGS868i Using PanaView™
Slot 1
(refer to Figure 55 on
119