Siemens MOBILETT XP Digital Troubleshooting Manual page 23

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

Error messages of the imaging system
Code
-3
-120
Siemens AG
Medical Solutions
Reply
Message
keys
OK
#1:Error
#2:Insufficient queue disk space. Output the saved data or
erase the backed up data.
#3:
[Cause]
The remaining space in the queue storage directory dtque is
less than the specified amount. Usually, this error does not
occur because the "Insufficient tmp disk space" error occurs
first.
[Operation after the reply key is pressed]
Press OK after confirmation.
OK
#1:Error
#2:Sensor unit: Self-check error
#3:Detected an error.#4:ER(n) (xxxx) n: Sensor index (1 to 4)
xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D board.
[Operation after the reply key is pressed]
Nothing
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 23 of 68
CS SD 24
23

Hide quick links: