62
•
Select "SERVICE TOOL > Explorer" in the menu.
•
Start via "START".
•
Insert the most recent labeled "Backup of the CONFIGURATION" CD into the CD drive.
•
Use Windows Explorer to copy the "e:\Configuration" folder to drive d:\TEMP.
•
Close Windows Explorer.
Restore of configuration files
•
Select "SYSTEM > CONFIGURATION".
•
Select "ADMINISTRATOR SETUP > SERVICE TOOL".
•
Click on the password box
•
Select "SERVICE TOOL > restore configuration to ccr-folder" in the menu.
•
Start via "START".
•
Close the window via "OK".
•
Exit the "SERVICE TOOL" window via "Exit".
•
Exit "ADMINISTRATOR SETUP" via "OK".
•
Switch off the MOBILETT XP Digital
•
Reinstall the PC and the top cover.
•
Restart the system.
MOBILETT XP Digital
Enter the Siemens service password and confirm with "OK" (refer to the pass-
word list).
- Enter the administrator password and confirm with "OK" (refer to the password
list).
Enter the Siemens service password and confirm with "OK" (refer to the pass-
word list).
All required files are copied to the D:\CCR directory.
SPR8-230.840.30.03.02
12.05
Backup and restore
Page 62 of 68
CS SD 24
Siemens AG
Medical Solutions