Checking The Firmware Of The Imaging System - Siemens MOBILETT XP Digital Troubleshooting Manual

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

Miscellaneous

Checking the firmware of the imaging system

Start up the MOBILETT XP Digital.
Select "SYSTEM > SETUP MENU > VERSION INFORMATION"
Fig. 10: CXDI firmware
-
A. Hardware version
This is the version of the detector hardware. It changes according to the setting of
the dipswitch on the PCB-50Di board in the detector.
-
B. Firmware initialization code version
This is the version of the initialization code written on the 50Di PCB board of the
detector. The initialization code will be downloaded and the settings will be reset to
the default (factory) settings by turning ON the power while pressing the initializa-
tion switch on the power box. Display "1.1.0" on the screen indicates version
1.01.00.
-
C. Firmware normal code version
This is the version of the normal code installed on the 50Di PCB of the detector.
Usually the system operates with this code. It must be updated as required. Dis-
play "1.1.00" on the screen indicates version 1.01.00.
Siemens AG
Medical Solutions
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 67 of 68
CS SD 24
67