Siemens MOBILETT XP Digital Troubleshooting Manual page 46

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

46
Code
-516
-517
-518
MOBILETT XP Digital
Error messages of the imaging system
Reply
Message
keys
RETRY
#1:Warning
CANCEL
#2:Internal storage error:
#3:Selected directory for storage does not exist, or the path is
incorrect.
[Cause]
There is no directory in the destination, or recording cannot be
performed due to a system function error.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored on the inter-
nal disk.When CANCEL is selected, the storing of the image will
be cancelled after a confirmation message is displayed.
Re-transfer is possible by sending the image from the study list.
RETRY
#1:Warning
CANCEL
#2:Internal storage error:
#3:Processed image cannot be stored, because the internal
storage memory capacity is insufficient.
[Cause]
The space on the internal disk is insufficient.
[Operation after the reply key is pressed]
The QA process will be performed again.
RETRY
#1:Warning
CANCEL
#2:Internal storage error:
#3:DICOM image cannot be stored, because the internal stor-
age memory capacity is insufficient.
[Cause] There is not enough space to store the DICOM image.
[Operation after the reply key is pressed] When RETRY is
selected, image will be stored on the internal disk.When CAN-
CEL is selected, the storing of the image will be cancelled after a
confirmation message is displayed.
SPR8-230.840.30.03.02
12.05
Page 46 of 68
CS SD 24
Siemens AG
Medical Solutions