Error messages of the imaging system
Code
-311
-402
Siemens AG
Medical Solutions
Reply
Message
keys
OK
#1:Error
#2:REX value warning
#3:REX is 9999. It is under the maximum limitation value 999.
Please reconfirm the generator setting or exposure condition. If
the condition is correct, adjust the image processing.
[Cause]
This error occurs when the REX value acquired by image pro-
cessing after imaging is more than the limit value.
[Operation after the reply key is pressed]
Nothing
[Note]
9999: REX value obtained by image processing 999: Upper limit
value
OK
#1:Error
#2:Incomplete end error
#3:The last image in the last study was not saved completely.
[Cause]
Failed to write the last image of the previous imaging operation.
[Operation after the reply key is pressed]
If previous images remain, they will be recovered.
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 29 of 68
CS SD 24
29