Error messages of the imaging system
Code
-302
-303
-304
Siemens AG
Medical Solutions
Reply
Message
keys
OK
#1:Error
#2:Calibration error
#3:Too much dose. Or collimator is used.
[Cause]
Excessive dose of radiation was used in exposure of image for
white correction.
[Operation after the reply key is pressed]
Perform imaging again.
OK
#1:Error
#2:Calibration error
#3:Too little dose. Or collimator is used.
[Cause]
Insufficient radiation dose was used in exposure of image for
white correction.
[Operation after the reply key is pressed]
Perform imaging again.
OK
#1:Error
#2:Calibration error
#3:Invalid data. Or collimator is used.
[Cause]
Too much difference in dose acquired by the panels.
[Operation after the reply key is pressed]
Perform imaging again.
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 27 of 68
CS SD 24
27