Error messages of the imaging system
Code
-153
-502
Siemens AG
Medical Solutions
Reply
Message
keys
OK
#1:Warning
#2:Sensor unit:Temperature warning
#3:The internal temperature of the sensor unit is approaching
the temperature limit. Do not leave the sensor unit in the READY
condition.
#4:TEMP OVER (n)(eff9) n: Sensor index (1 to 4)
[Cause]
The warning sign is detected because the internal temperature
of the imaging unit has approached the upper limit value.
[Operation after the reply key is pressed]
The sensor automatically enters the state of sleep. No operation
is necessary on the host side. The same error occurs when the
user tries to put this sensor into the READY state. In this case,
press [OK] to close the window and continue to operate. This
error might be able to be avoided by reducing the internal tem-
perature or shortening the managing unit READY STATE TIME.
RETRY
#1:Warning
CANCEL
#2:DICOM connect error
#3:Cannot connect to target. Check network or port number set-
ting.
[Cause] Cannot be connected to the destination.
[Operation after the reply key is pressed] When RETRY is
selected, the image will be transferred again. When CANCEL is
selected, transfer will be canceled after a confirmation message
is displayed. Re-transfer is possible by sending the image from
the study list.