Siemens MOBILETT XP Digital Troubleshooting Manual page 48

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

48
Tab. 8
Customer Warning Table
Error Code
-150
-151
-152
-153
-160
-161
-162
-163
-164
-502
-503
-504
-505
MOBILETT XP Digital
Error messages of the imaging system
Error Message
Detector Unit: Self Check Error:
Detected an error.
Detector Unit: Flash ROM Error:
Detected Flash ROM Error. Call
a service person.
Detector Unit: Sequence Error:
This image may have a problem.
Detector Unit: Temperature
Warning: The internal tempera-
ture of the detector unit is too
high! Do not leave the detector
unit in READY condition.
Capture Board Error: Detected
capture board error.
Capture Board Error: Detected
capture board error. Call a ser-
vice person.
Capture Board: Sequence Error:
X-ray monitor in detector unit
failed to detect start of radiation.
Capture Board: Sequence Error:
X-ray monitor in detector unit
failed to detect end of radiation.
Image Quality attention: FD has
stopped radiation. Check image.
DICOM Connect Error: Cannot
connect to target. Check net-
work or port number setting.
DICOM Transfer Error: Error has
occurred during association.
DICOM Parameter Error:
DICOM transfer parameter is
wrong.
DICOM Response Error: Receiv-
ing device reported an error in
DICOM response. Check receiv-
ing device.
SPR8-230.840.30.03.02
12.05
Remedy
Touch OK and shut down the sys-
tem.
If these errors occur, contact a Sie-
mens representative or distributor.
Touch OK and repeat the exposure,
since the image quality may be de-
graded.
Touch OK to close the dialog win-
dow.
Lower the temperature of the room
where the detector unit is installed,
or wait until the temperature in the
detector unit drops.
Touch OK and shut down the sys-
tem.
If these errors occur, contact a Sie-
mens representative or distributor.
Touch OK and repeat the exposure,
since the image quality may be de-
graded.
After confirming that the network is
OK, touch RETRY. Or touch Cancel
and check the settings.
Touch RETRY. Or touch Cancel
and check parameter settings.
Check that the printer or storage me-
dium is OK. Then touch RETRY.
Or touch Cancel and check the DI-
COM settings.
Page 48 of 68
CS SD 24