MOBILETT XP Digital
Error messages of the imaging system
Error Message
Detector Unit: Self Check Error:
Detected an error.
Detector Unit: Flash ROM Error:
Detected Flash ROM Error. Call
a service person.
Detector Unit: Sequence Error:
This image may have a problem.
Detector Unit: Temperature
Warning: The internal tempera-
ture of the detector unit is too
high! Do not leave the detector
unit in READY condition.
Capture Board Error: Detected
capture board error.
Capture Board Error: Detected
capture board error. Call a ser-
vice person.
Capture Board: Sequence Error:
X-ray monitor in detector unit
failed to detect start of radiation.
Capture Board: Sequence Error:
X-ray monitor in detector unit
failed to detect end of radiation.
Image Quality attention: FD has
stopped radiation. Check image.
DICOM Connect Error: Cannot
connect to target. Check net-
work or port number setting.
DICOM Transfer Error: Error has
occurred during association.
DICOM Parameter Error:
DICOM transfer parameter is
wrong.
DICOM Response Error: Receiv-
ing device reported an error in
DICOM response. Check receiv-
ing device.
SPR8-230.840.30.03.02
12.05
Remedy
•
Touch OK and shut down the sys-
tem.
•
If these errors occur, contact a Sie-
mens representative or distributor.
•
Touch OK and repeat the exposure,
since the image quality may be de-
graded.
•
Touch OK to close the dialog win-
dow.
Lower the temperature of the room
where the detector unit is installed,
or wait until the temperature in the
detector unit drops.
•
Touch OK and shut down the sys-
tem.
•
If these errors occur, contact a Sie-
mens representative or distributor.
•
Touch OK and repeat the exposure,
since the image quality may be de-
graded.
•
After confirming that the network is
OK, touch RETRY. Or touch Cancel
and check the settings.
•
Touch RETRY. Or touch Cancel
and check parameter settings.
•
Check that the printer or storage me-
dium is OK. Then touch RETRY.
•
Or touch Cancel and check the DI-
COM settings.