Siemens MOBILETT XP Digital Troubleshooting Manual page 33

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

Error messages of the imaging system
Code
-520
-521
-599
Siemens AG
Medical Solutions
Reply
Message
keys
OK
#1:Error
#2:Internal storage error:
#3:Exposed image cannot be stored, because the internal stor-
age memory capacity is insufficient.
[Cause]
The remaining space on the internal disk is insufficient.
[Operation after the reply key is pressed]
The system returns to the state before the occurrence of the
error.
OK
#1:Error
#2:Queue file open error:
#3:Queue file is not found.
[Cause]
Failed to open the queue file. This error occurs when no queue
file exists.
[Operation after the reply key is pressed]
*With the current system, this error is unlikely to occur.
OK
#1:Error
#2:File open error:
#3:File is not found.
[Cause]
This error occurs when the system has failed to open the file
during RESEND.
[Operation after the reply key is pressed]
Simply press OK.The system returns to the state before the
occurrence of the error.
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 33 of 68
CS SD 24
33