Siemens MOBILETT XP Troubleshooting Manual page 28

Hide thumbs Also See for MOBILETT XP:
Table of Contents

Advertisement

28
Test points D916 (CPU)
SP
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
MOBILETT XP
Name
Function / information
GND
0V signal (ground)
GND
0V signal (ground)
KVN
kV actual, negative
KVP
kV actual, positive
KV
kV actual, total
JR
mA actual, tube current (mA)
I
Primary resonant filament current (can be mea-
sured only with the oscilloscope)
IH
Actual effective filament current
SWT
Main inverter control circuit (high = kV actual >
70%)
F1
mAs integration
TEMP
Single tank temperature (analog/neg. coefficient)
VR
+5 V - DC, U1
VC
Capacitor bank voltage
SCM
Main inverter short-circuit (high = error, e.g.: ERR
22)
SCH
Filament inverter short-circuit (high = error, e.g.:
ERR 36)
JIM
Overcurrent error (filament or tube current), (high
= error, e.g.: ERR 23/ERR 42)
KVM
Overvoltage error (+/- kV), (high = error, e.g.:
ERR 31)
KVC
Main inverter resonant current
VSA
Not used
KVS
kV nominal (kV)
JRS
Nominal tube current (mA)
IHS
Nominal effective filament current
I1
Motor current (stator winding 1)
I2
Motor current (stator winding 2)
FC1
Filament inverter (kHz)
FC2
Main inverter (kHz)
SPR8-230.840.01.05.02
11.05
Test Points, Fuses and LEDs
Page 28 of 72
CS SD 24
Test ratio
n.a.
n.a.
1 V ^ 10 kV
1 V ^ 10 kV
1 V ^ 30 kV
1 V ^ 100 mA
1 V ^ 1 A
1 V ^ 2 A
n.a.
100mA ^ 16 kHz
> 2 V = > 60°C
± 0,01
1 V ^ 100 V
n.a.
n.a.
n.a.
n.a.
1 V ^ 50 A
n.a.
1 V ^ 30 kV
1 V ^ 100 mA
1 V ^ 2 A
1 V ^ 4 A
1 V ^ 4 A
n.a.
n.a.
Siemens AG
Medical Solutions

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett xp ecoMobilett xp hybridMobilett xp digital

Table of Contents