Siemens MOBILETT XP Troubleshooting Manual page 45

Hide thumbs Also See for MOBILETT XP:
Table of Contents

Advertisement

Error messages
ERR 92
System
status:
Initializa-
tion
ERR 93
System
status:
Initializa-
tion
ERR 96
System
status:
Initializa-
tion
ERR 97
System
status:
Initializa-
tion
ERR 98
System
status:
Initializa-
tion
Siemens AG
Medical Solutions
RAM error
Explanation: D916 (CPU) detects a RAM error.
Action:
Check the supply voltages of power supply U1.
Check fuses F7 and F8 on D927, relay K22 and the cable connec-
tions.
Check the voltages at D916 X15.
- +15 V D916 X15 pin 5-6
- -15 V D916 X15 pin 6-7
- +5 V D916 X15 pin 4-3
Voltage error of +15 Vcc
Explanation: D916 (CPU) detects a voltage error of +15 Vcc.
Action:
Check the voltages of power supply U1 and D916 X15.
- +15 V D916 X15 pin 5-6
- -15 V D916 X15 pin 6-7
- +5 V D916 X15 pin 4-3
kV converter error
Explanation: CPU board D916 is defective.
mA converter error
Explanation: CPU board D916 is defective.
I
converter error
H
Explanation: CPU board D916 may be defective.