User message 01
Explanation: If the exposure time from the digital application software
(parameter kV, mAs, mA) is < 1 sec but the calculated exposure time of the
generator is > 1 sec, the USE 01 message is displayed.
Action:
Due to a possible faulty exposure, the mA value in the digital application
software is to be reduced.
Error value +15 V
Explanation: D916 (CPU) faulty recording of +15 V DC. System cannot be
switched on.
Action:
Check D916 (CPU). X15.5-6 voltage "+15 V" from power supply U1.
Oil pressure too high or tube too hot
Explanation: Oil pressure switch or temperature sensor of the single tank
activated.
Action:
Allow single tank to cool prior to check.
Check the connections and cable between D916 (CPU X8.9/X8.8 and oil
pressure switch/temperature switch on the single tank (K2).
Static short-circuit at main inverter D962
Explanation: D916 (CPU) detected main inverter short-circuit (D962).
Action:
Check D962: Switch system off, wait 15 minutes (capacitor discharge
time!), disconnect the cable connection at C1, switch the system on
again.
- If ERR 4 is displayed again, D962 must be replaced.
- If ERR 4 is not displayed again, the problem is in the single tank, ca-
pacitor C1, or the cable connection/plug to the single tank.
SPR8-230.840.01.05.02
11.05